BenchCAT 4000R HP Hybrid

Multi-station Hybrid Reactor and Chemisorption: Chemi-CAT
The BenchCAT 4000R HP Hybrid is a four station automated reactor system designed to:
- Conduct temperature-programmed or isothermal experiments with catalysts loaded in a straight tube reactor at ambient or pressure conditions. External analysis will be performed with a mass spectrometer.
- Perform characterization experiments (TPR, TPO, TPD, Pulse Chemisorption) using catalysts at ambient pressures. Analysis will be made internally, using the TCD, or externally with a mass spectrometer, GC, etc….
- The instrument is configured to perform either reaction experiments or characterization experiments on an automated basis. Reaction and characterization will always be performed as separate experiments.
78cm x Reaction Testing Specifications
Number of Stations: 4
System Operating Pressure Ambient: 100 barg (higher/lower pressures available)
MFC Inlet Pressure: (Varies according to operating pressure)
Maximum Furnace Operating Temperature: 1200°C (depends on pressure and reactor material)
Maximum Post Reactor Line Temperature: 150°C
Number of MFCs (Reaction Side): Determined by the customer
Number of MFCs (Shared): Determined by the customer
Number of Vaporized Liquid Feeds: Determined by the customer
Materials of Construction
Upstream Plumbing: 316SS
Downstream Plumbing: 316SS
Sample tubes 316SS, Quartz, Inconel or otherwise determined by the operating conditions
Sample tube: option monolith, powder, or pellets
Wetted Parts: 316SS
Seal Materials: Premium
Catalyst Charge: Determined by the customer
Overall System Footprint 178cm H x 153cm W x 77cm L (Approx)
Utility Requirements
Oil-free, Dry Air: Multiple 80 psig utility connections
System power: 220 VAC
Computer power: 110V or 220V
Resources & Downloads
BenchCAT 4000R HP Hybrid Enquiry
For more information or to arrange a call from one of the Crea Particle Science Team, please complete the contact form . If your enquiry is urgent, please call us on 1300 552 725.
Catalyst Characterization
Number of Stations 4
System Operating Pressure Ambient
Gas Inlet Pressure Range 50 psig
Gas Outlet Pressure Range near ambient
Maximum Furnace Operating Temperature 1200°C
Maximum TCD Operating Temperature 200°C
Materials of Construction
Plumbing 316SS
Sample U-tubes Quartz
Wetted Parts 316SS
Seal Materials Premium seals
Catalyst Charge 0.1 – 1g
Number of Treatment/Carrier Ports 4 (per station)
Number of Blend Ports 2 (per station)